Customer Resources

Quality Management

Western States Calibration operates under the requirements of ISO 17025, ISO 9001, and ANSI/NCSL Z540-1 with measuring standards traceable to the National Institute of Standards and Technology.

ISO 17025:2017 Decision Rules:

By default, Western States Calibration (WSC) provides statements of conformity by applying a simple 1:1 decision rule per ASME B89.7.3.1. In applying this decision rule the TUR (test uncertainty ratio) shall be equal to or greater than 1:1. WSC’s default decision rule results in a worst-case estimated Probability of False Accept (PFA) of <2%.

If you require the use of some other decision rule, please contact us with your request.

Note: Custom decision rules may result in a price increase for your calibration services.

Calibration Management System


A2LA Scope of Accreditation

A2LA Scope Listing


Equipment Background Form

This form should be completed the first time an instrument is calibrated by WSC at your facility, or if an instrument has special requirements to be tested.


Equipment Order Information Form

This form should be completed and included with all orders submitted to WSC.


QPM (Permission Control)

To request a copy of our QPM, contact [email protected]